RATAN-600 Radiometer = 2.7 cm
Technical information:
Central frequency
11.2 GHz
Bandwidth
1.2 GHz
Bandpass Filter
10.7 - 11.9 GHz
Front end amplifier
ξεντ
Physical temperature
15 K
System noise temperature
66 K
Realized sensitivity ( = 1 s )
3 mK
Design
Switching Radiometer
Operating modes
Double - beam switching or
Single - beam mode with a cooled reference
Polarization
No
Feed Horn
Quasi Scalar
Calibration
Noise generator

Simplified block-diagram of the radiometer

  • PF1, PF2 - input feed horns,
  • DC1, DC2 - directional couplers,
  • S1, S2 - latching circulator switches,
  • CI - cooled ferrite isolator,
  • GM - p-i-n diode gain modulator,
  • BPF - band-pass filter,
  • D - square law detector,
  • NG1 - noise generator for calibration,
  • NG2 - balance noise generator,
  • A1 - noise power control attenuator,
  • LFA - low frequency amplifier.
Operating modes:

1. For double-beam mode of operation both latching circulator S1 and S2 are switched in phase. The input feed horns H1 and H2 are connected alternatelly to HEMT amplifier.
2. For single-beam mode latching circulators S1 and S2 are switched in antiphase. The cooled matched load operates as a reference. The balance is realized by means of noise adding to the matched load.

Interior of the 15 K cryostat:


  • 1 - cooled modulators ( latching circulator switches ),
  • 2, 8 - thin-wall thermo-insulating waveguides,
  • 3 - waveguide ferrite isolator,
  • 4 - waveguide-to-coaxial adapter,
  • 5 - HEMT amplifier,
  • 6 - output coaxial ferrite isolator,
  • 7 - cooled matched load,
  • 9 - anti-radiating shield,